[HPLC] High Performance Liquid Chromatography

HPLC:High Performance Liquid Chromatography

HPLC equipment

Features

High-performance liquid chromatography (HPLC) is a chromatographic method that separates and identifies components in liquids.
By selecting separation methods and detectors with respect to the target sample, qualitative and quantitative determination of organic compounds of low and high molecular weight as well as low and high polarity, is possible.

  • Compatible with highly polar compounds of low molecular weight such as organic acids as well as with low-polar compounds such as lipids
  • Impact of impurities can be eliminated by the post-column method, even in the case when contaminants cannot be separated
  • Determination of molecular weight and molecular weight distribution of polymers

Application Examples

  • Quantitative analysis of functional ingredients in foods
  • Quantitative analysis of organic and amino acids in food and industrial products
  • Quantitative analysis of lipids and fatty acids
  • Quantitative analysis of monosaccharides and polysaccharides
  • Quantitative analysis of polymers
  • Molecular weight distribution measurement of polymers

Principle of operation

The first step is column separation in which the liquid (or dissolved) sample is introduced into the column. Each component of the sample is then separated by the difference in affinity (retention force) between the stationary phase and the mobile phase in the column.
HPLC allows the use of low-polarity solvents, such as hexane and chloroform, but also polar solvents, such as phosphoric or acetic acid buffers, water, and acetonitrile, in the mobile phase. Therefore, various separation methods can be selected by combining them in several different columns in tandem.
After separation, both absorption (Fig. 2) and fluorescence (Fig. 3) spectra can be measured to obtain detailed information on the substance.

Absorption spectrum measurement using photo-diode array (PDA) detector covering the ultraviolet-visible range (190-800nm).
Fluorescence spectrum measurement using a Xe lamp as excitation source, a monochromator , and a high-sensitivity photo multiplier tube (PMT).

Additional information can be obtained from scattering measurements (Fig. 4), for example, detection of substances that cannot be identified by absorption or fluorescent spectroscopy. The liquid exiting the column (eluate) is sprayed with a nebulizer using an inert gas to form small droplets the solvent of which is subsequently evaporated by heating. The vapour of remaining particulates are irradiated with 400-900 nm focused light and the amount of scattering is detected.

Evaporative Light Scattering (ELS) measurement.

Also, the change in refractive index by the sample can be measured (Fig. 5) to gain further information about the substance.

Measurement of refractive index (RI) change via photorefraction between the reference and a test cell. When the RI changes, the angle of the reflected light changes. From this change, the presence of a substance can be confirmed.

Data examples

Fig. 6 shows an example of a chromatogram of aldehydes derivatized by 2,4- dinitrophenylhydrazine (DNPH), or also known as Brady's reagent. The concentration by weight of each substance can be calculated from the peak areas.

Chromatogram showing the relative amount of absorption at 360 nm.

The chromatogram in Fig. 6 can also be combined with absorption spectra (see Fig. 7). The amount of absorption at characteristic wavelengths corresponds to the amount of each component.

Spectral chromatogram of aldehydes derivatized by DNPH (absorption wavelength range: 210-400 nm). The inset shows the absorption spectra for the substance with a hold time of 0.5 minutes (formaldehyde).

Data delivery format

  • Portable Document Format (PDF) file
  • Numerical data of the chromatogram in .csv format

Specifications

Property

Items for enquiries

  1. Purpose and content of measurement
  2. Sample information
    ・ Number of samples, and availability of preliminary samples
    ・ Sample type (thin film, powder, solution)
    ・ Solvents that can be dissolved if solution adjustment is required
    ・ Handling instructions
  3. Details on delivery
    ・ 全ての情報はこちら Preferred due date for preliminary analysis report
    ・ Due date for delivery of final report
    ・ Priority in the case of a large number of samples
  4. Any other issues

Caution

  • Analysis may be impossible due to significant effects of impurities and other components that can interfere with the measurement.

[HPLC]高速液体クロマトグラフ法の分析事例はこちらからご覧ください。

Consultation and application for analysis

Our knowledgeable sales representatives will propose the most appropriate analysis plan.
Please feel free to contact us for a quote on the cost of your analysis.
For consultation and application, please use the inquiry form or call us.

Inquiry Form

Contact Us

TEL

ToPageTop

てむぞう&ますみん

Temuzo&Masumin